LUTRON PCD-431 CONDUCTIVITY METER HANDY
Overview
The Lutron PCD431 is an economical yet rugged pocket tester for conductivity and TDS. It is autoranging with selectable TDS or conductivity, and so replaces up to 4 different models from our main competitor. Accuracy is +/-3% of fullscale. Because of the adjustable ATC factor, good accuracy is possible over a wide temperature range. Auto power off and rugged IP67 and sensor cover makes this a very handy mobile tool.
FEATURES
* All in one pen type conductivity meter provides fast,
accurate readings with digital reading.
* Conductivity measurement ( uS, mS ) or TDS
( Total Dissolved Solids, ppm ) can be selected.
* Conductivity : Two ranges, 2,000 uS, 20.00 mS.
* TDS : Two ranges, 20,000 ppm, 2,000 ppm.
* Carbon rod electrode for long life.
* Build in temperature sensor, ATC ( auto temperature
compensation ).
* Temperature measurement, ℃, ℉.
* IP67, water proof and protection.
* LCD with two displays show conductivity and Temp value at
same time.
* Data hold function for freezing the desired value.
* Auto power off to save the battery life.
* Records max. and min. value with recall.
* Microcomputer circuit, intelligent function, high accuracy.
* Compact size, light weight.
* Power supply by DC 1.5 V battery ( UM4/AAA ) x 4 PCs,
* Available for wide applications, such as aquarium, beverage,
fish hatcheries, food processing, photography, laboratory, quality
control, school & colleges, swimming pools, water conditions.
SPECIFICATIONS
Display
LCD, size : 20 mm x 28 mm.
Measurement
* Conductivity ( uS, mS )
* TDS ( Total Dissolved Solids, ppm )
* Temperature ( ℃,℉)
Ranges
* two ranges
* auto range
Conductivity: 2000 uS, 20.00 mS
TDS : 2,000 ppm 20,000 ppm
Accuracy
* 23 ± 5 ℃
± ( 2% FS + 1 d )
* FS : full scale
Temperature
Compensation
Automatic from 0 to 60 ℃ (32 - 140 ℉),
with temperature compensation factor variable between 0 to 5.0% per C.
Conductivity
Probe Structure
Carbon rod electrode for long life.
Data Hold
Freeze the display reading.
Memory Recall
Maximum & Minimum value.
Sampling Time
Approx. 0.8 second.
Circuit
Custom one-chip of microprocessor LSI circuit.
Power off
Auto shut off saves battery life or manual off by push button.
* Power will off automatically after 10 min., if
no button be pressed.
Operating Temperature
0 to 50 ℃ ( 32 to 140 ℉ ).
Operating Humidity
Less than 80% RH.
Power Supply
DC 1.5V battery ( UM-4/AAA ) x 4 PCs.
Power Consumption
Approx. 5.7 mA.
Dimension
190 x 40 x 40 mm ( 7.5 x 1.6 x 1.6 inch ).
Weight
171 g/0.38 LB.
Standard Accessories
Instruction Manual........................... 1 PC
Power off
Auto shut off saves battery life or manual off by push button.
* Power will off automatically after 10 min., if no
button be pressed.
Standard Accessories
Instruction Manual........................... 1 PC
Optional Accessories
* Soft carrying case with sash
( 210 x 80 x 50 mm ), Model : CA-52A
* Hard carrying case
( 280 x 195 x 65 mm ), Model : CA-06
* 1.413 mS Conductivity Standard Solution,
Model : CD-14A
ELECTRICAL SPECIFICATIONS (23± 5 ℃)
A. Conductivity
Range
Measurement
Resolution
Accuracy
2000 uS
0 to 2000 uS
1 uS
± (3% F.S.+1d)
* F.S. - Full scale
20 mS
2 to 20.00 mS
0.01 mS
* Auto range.
* Temperature Compensation :
Automatic from 0 to 60 ℃ ( 32 - 140 ℉ ), with temperature
compensation factor variable between 0 to 5.0% per C.
* mS - milli Simens
* uS - micro Simens
* 23± 5℃
B. TDS ( Total Dissolved Solids )
Range
Measurement
Resolution
Accuracy
2,000 ppm
0 to 1,320 ppm
1 ppm
± (3% F.S.+1d)
* F.S. - Full scale
20,000 ppm
1,320 to 13,200 ppm
10 ppm
* Auto range.
* Temperature Compensation :
Automatic from 0 to 60 ℃ ( 32 - 140 ℉ ), with temperature
compensation factor variable between 0 to 5.0% per ℃.
* ppm - parts per million
* 23± 5℃
C. Temperature
Function
Measuring Range
Resolution
Accuracy
℃
0 ℃ to 60 ℃
0.1 ℃
0.8 ℃
℉
32 ℉ to 140 ℉
0.1 ℉
1.5 ℉