Specifications
Capacity(g) x Readability (mg):0-40 g x 0.01 mg | 40-60 g x 0.01 mg | 60-120 g x 0.1 mg
Reproducibility (std deviation) milligrams (mg):± 0.03 mg [0-40g] | ± 0.04 mg [40-60g] | ± 0.07 mg [60-120g]
Linearity milligrams (mg):± 0.1 mg [0-40g] | ± 0.1 mg [40-60g] | ± 0.2 mg [60-120g]]
Response Time (average):6 seconds [0-60g] | 2 seconds [60-120g]
Tare Range:FULL to Capacity
Selectable Weight Units:g, ct, mom, kg, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, parts | lb, tlc, Kt, tol, bat, MS, N
Selectable Application Programs:Counting, Weighing in percent, Density determination, Peak hold, Unstable condition, Checkweighing, Mixing, Conversion, Statistics, Components.
Languages:English, German, French, Spanish, Italian, Russian, Chinese, Polish, Japanese, Portuguese
Display:Touch screen with Sartorius graphical user interface
Allowable Ambient Operating Temperature:10° C to 30° C (50° F to 86° F)
Sensitivity Drift(10° - 30°C/50°F-86°F):± 1.5 ppm / °K
Motorized Internal Calibration:YES
External Calibration:100g Optional, not included
Pan Size:ø80 mm / 3.1 in
Draft Shield Chamber Height(from pan to top of glass door):218 mm / 8.58 in
Dimensions (entire balance) W x D x H:218 x 359 x 319 mm | 8.6 x 14.1 x 12.6 in
Interface:mini USB
Weighing Below:Yes
Minimum sample weight acc. to USP (typical):Not Available only with Secura Series
Leveling:Manual with front level bubble with two front leveling feet
GLP/GMP Data Capability:YES GLP printout capability
Internal Clock:Yes
AC Power Source:AC Adapter 120 VAC
Housing Design:Aluminum Die-cast base with coated ABS top housing
Weigh Sensor:Magnetic Force Restoration
Place of Origin:Germany
Standard Equipment:120V AC Adapter, In Use Dust Cover, Manual
Net Weight, Approx.:8.6