Ventilation and Control
--natural convection
--continuous adjustment of pre-heated fresh air admixture
--vent connection with restrictor flap
--fuzzy-supported PID microprocessor controller with integrated auto diagnostic system with fault indication
--solid state switching unit
--2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
--digital 7-day-programme-timer with real time clock, precise minute setting, for one set value or ramp operation in combination with:
--integrated digital timer for tempering profile with max. 4 segments, each segment adjustable from 1 min to 999 hours: DELAYED ON ,HEATING UP
, HOLD or HOLD set-temperature-dependent with guaranteed holding time, COOLING DOWN
--"Loop" function to repeat profile 1-99 times or endless
--digital display (LED) of all set parameters, such as temperature, weekdays, time, programme status and set-up values
--resolution of display for set value and actual value 0,1 °C
--long-term logging (ring store) of all relevant data, GLP-conforming as data logger - 1024kB
--serial interface RS 232 including MEMMERT Software "Celsius" for programming and documentation
--incl. works calibration certificate for +37 °C