On this page, you can find all the essential technical data on the Memmert water bath. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Temperature
Setting temperature range
+10 to +95 °C and boiling stage
Working temperature range in °C
min. 5 above ambient up to +95 °C with additional boiling mode
Control of standard components
Temperature sensor
1 Pt100 sensor class A in 4-wire-circuit
Timer
integrated digital timer from 1 min. to 99,59 hours for: ON continuous operation WAIT (delayed on for continuous and limited timed operation) HOLD
Control technology
Controller
digital display of set and actual temperature and of (remaining) programme time
Safety
Temperature control
mechanical temperature limiter TB protection class 1 switching the heating off at approx. 30°C above max. temperature of the bath
Temperature control
in case of overtemperature due to failure, the heating is switched off at approx. 10°C above the set temperature (fixed value)
Autodiagnostic system
microprocessor PID-temperature controller with integrated autodiagnostic system with fault indicator
Heating concept
Heating Baths
corrosion-proof large-area heating on three sides
Stainless steel interior
Dimensions W x H x D in mm
w(A) x h(C) x d(B): 350 x 140 x 290 mm
Interior
easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing, material no. 1.4301 (ASTM 304), laser-welded
Volume
14 l
Liquid level min.
97 mm
Liquid level max.
120 mm
Textured stainless steel casing
Dimensions
w(D) x h(G) x d(E): 578 x 347 x 436 mm
Electrical data
Voltage
230 V, 50/60 Hz
Electrical load
approx. 1800 W (during heating)
Ambient conditions
Installation
The vent openings in the left and back side must remain unobstructed. Minimum wall spacing on all sides is 80 mm. The minimum spacing from the top of the bath to the next ceiling is 750 mm.